. 2011 Oct 13;5:61. doi: 10.1186/1752-153X-5-61

Table 1.

Performance data of the proposed method.

Parameter Spectrofluorimetric Method Spectrophotometric Method Ref. method [23]
-concentration range (μg/mL ). 0.25-5 2-24
-LOD (μg/mL ). 0.069 0.184
-LOQ (μg/mL). 0.209 0.558
-Correlation coefficient ( r). 0.9999 0.9999
-Slope 31.8 0.035
-Intercept 19.038 0.039
-Sy/x 0.928 2.72 × 10-3
-Sa 0.664 1.96 × 10-3
-Sb 0.221 1.387 × 10-4
-% Error 0.29 0.21
-%RSD 0.65 0.51
-No.of Experiments. 5 6 3
-Mean found (%) 100.12 100.74 100
± SD. 0.65 0.51 0.31
-Student's t-value. 0.275 (2.45) 2.216 (2.37)
-Variance ratio F-test. 4.352 (19.25) 2.762 (19.30)
-Applications. tablet preparations and spiked human plasma

N.B.

-Sy/x =standard deviation of the residuals.

-Sa = standard deviation of the intercept of regression line.

-Sb = standard deviation of the slope of regression line

-% Error = RSD%/√ n.

- Figures between parentheses are the tabulated t and F values respectively, at p = 0.05.[25]