Table 1.

Results for the Omron M2 device according to the ESH international protocol

Phase 1 ≤5 mmHg ≤10 mmHg ≤15 mmHg Recommendation
Required One of 25 35 40
Achieved SBP 23 38 42 Continue
DBP 40 45 45 Continue
Phase 2.1 5 mmHg 10 mmHg 15 mmHg Recommendation Mean difference (mmHg) SD

Required Two of 65 80 95
All of 60 75 90
Achieved SBP 73 92 96 Pass 2.7 5.0
DBP 88 99 99 Pass −1.4 3.2
Phase 2.2 2/35 mmHg 0/35 mmHg Recommendation

Required ≥22 ≤3
Achieved SBP 25 3 Pass
DBP 31 0 Pass

Abbreviations: ESH, European Society for Hypertension; SD, standard deviation (mmHg); SDP, systolic blood pressure; DBP, diastolic blood pressure.