Skip to main content
. Author manuscript; available in PMC: 2013 Jan 1.
Published in final edited form as: J Heart Lung Transplant. 2011 Oct 20;31(1):101–105. doi: 10.1016/j.healun.2011.08.022

Table 2.

Summary of the in-vivo animal experiments (PFH data are given as mean ± standard deviation.)

Sheep
Number
Weight
(kg)
Duration
(hour)
Device Flow
(L/min)
02 Transfer
(ml/min)
PFH
(mg/dL)
Plasma
Leakage
1 50.0 6 1–4 63–173 10.7±3.6 no
2 46.5 24 0.5–4 33–188 7.2±2.6 no
3 50.9 24 2.9–3.4 114–185 12.5±3.3 no
4 53.8 24 2.9–3.2 89–172 13.6±3.6 no