Table 1.

Comparison of typical characteristics for thermal and photochemical damage

Property Thermal Damage Photochemical Damage
Exposure duration µs – 10 s > 1 s
Wavelength all λ short visible (λ < 600 nm)
Time course appear in < 24 hrs appear in 24 – 48 hrs
Temperature change > 20 °C temperature ↑ < 10 °C temperature ↑
Minimal lesion size < beam diameter = beam diameter
Reciprocity of time & power no yes
Scotoma permanent reparable
Damage threshold ∝ (irradiated area diameter)−1
∝ time−1/4
independent of λ (for 440 nm < λ < 700 nm)
independent of irradiated area
∝ time−1
∝ λ