Iron sucrose removal in a closed-loop system by high-flux or high-efficiency hemodialysis membranes over 240 minutes. (System B)

Iron Sucrose (Venofer®)

Study condition SBS % iron concentration change
(mean ± SD)
Khd (hr-1)
(mean ± SD)
Dialysate Iron Recovery (mg)
(mean ± SD)
Clearance(ml/min)
(mean ± SD)
F80: 0 ml/hr UFR -3.13 ± 0.06 -0.0001 ± 0.0000 13.82 ± 13.34 -0.76 ± 0.01
F80: 500 ml/hr UFR -0.03 ± 1.15 0.0000 ± 0.0000 23.59 ± 3.96 0.05 ± 0.19
CA210: 0 ml/hr UFR -1.41 ± 3.77 -0.0001 ± 0.0002 5.47 ± 1.52 -0.34 ± 0.92
CA210: 500 ml/hr UFR -1.58 ± 3.08 -0.0003 ± 0.0002 2.79 ± 1.89 -1.08 ± 0.93

SBS = simulated blood solution; UFR = ultrafiltration rate; Khd = hemodialysis elimination rate; % = percent; † = Clearance values that are positive relate to accumulation of substance from the system; negative clearance values represent removal of substance