Table 1.

Imaging protocol with sequence specifications. Volumetric images were utilized to identify anatomical landmarks used to quantify target displacement. * The initial two volumetric images (2/13) were gradient echo sequences (first 2 procedures using the robotic technique)

Sequence TR/TE/FA ms/ms/degrees Resolution (mm) Acquisition time (minutes)
Diagnostic multi-parametric image sequences
 DWI, b-values: 50, 500, and 800 s/mm2. 2300–2500/61–64 2.0 × 2.0 × 4.0 3:08
 Transversal, sagittal and coronal T2-w turbo spin echo 4480–4950/103–110/120 0.6 × 0.6 × 3.0 3:22–4:43
 DCE-MRI 3D T1-w spoiled gradient-echo 32/1.47/10 1.8 × 1.8 × 4.0 2:43
Image sequences during biopsy procedure
 DWI, b-values: 0, 100, 500 and 800 s/mm2. 2000/67 1.8 × 1.8 × 4.0 2:06
 T2-w turbo spin echo 3620/103/120 0.8 × 0.8 × 4.0 3:26
 Transversal and sagittal TRUFI image (manual technique) 4.48/2.24/70 1.1 × 1.1 × 3.0 7.5 and 8.9 s
 Transversal, sagittal and coronal TRUFI image (robotic technique) 894/2.3/60 1.6 × 1.6 × 5.0 0.9 s/slice
 T1-w 3D volumetric gradient echo* 4.5/2.2/43 1.0 × 1.0 × 1.0 2:20
 T2-w 3D volumetric spin echo 1000/102/100 1.0 × 1.0 × 1.0 2:36