Skip to main content
. 2011 Nov 29;470(2):402–409. doi: 10.1007/s11999-011-2193-3

Table 1.

Test study design

Test Impaction test description Impact force (lbs) Number of strikes Sample size
1 Misaligned inserts (2–4 mm) a) 1000 2 n = 5
b) 2000 2
2 Deformed shells from 1-mm oblong underream a) 1000 2 n = 5
b) 2000 2
3 Taper with soft tissue impingement a) 1000 2 n = 5
b) 2000 2
4 Taper damage using a 0.25-mm diameter wire a) 1000 2 n = 5
b) 2000 2
5 Misaligned inserts (2–4 mm), deformed shells from 1-mm press-fit, taper damage using a 0.25-mm diameter wire a) 1000 2 n = 5
b) 2000 2
6 Misaligned inserts (2–4 mm) a) 2000 10 n = 5 (from Test 1)
b) 3000 10
c) 4000 10
d) 5000 10
7 Deformed shells from 1-mm oblong underream a) 2000 10 n = 5 (from Test 2)
b) 3000 10
c) 4000 10
d) 5000 10
8 Taper damage using a 0.25-mm diameter wire a) 2000 10 n = 5 (from Test 4)
b) 3000 10
c) 4000 10
d) 5000 10
9 Misaligned inserts (2–4 mm), deformed shells from 1-mm press-fit, taper damage using a 0.25-mm diameter wire a) 2000 10 n = 5 (from Test 5)
b) 3000 10
c) 4000 10
d) 5000 10