Skip to main content
. 2011 Nov 14;6:150. doi: 10.1186/1749-8090-6-150

Table 1.

Clinical data at implantation, weaning and post-explantation

Time/Parameter Pre ECMO 3 months post LVAD 9000 rpm rest 6000 rpm rest 9000 rpm exercise 6000 rpm exercise p-explant 9 months p-explant
NYHA IV I-II I-II I-II -- -- -- I-II

Hemodynamics

Systolic blood pressure, mm Hg 70 -- -- -- -- -- 80 90

Mean blood pressure, mm Hg 40 95 80 85 180 130 - -

Diastolic blood pressure, mm Hg - -- -- -- -- -- 50 60

Heart rate, beats/min 130 66 70 85 182 185 110 73

Cardiac index, L/min/m2 - - - 3.2 - 4.9 2.9 --

Pulmonary capillary wedge, mm Hg - - - 11 - 22 - --

Pulmonary artery systolic, mm Hg - - - 21 - 40 28 --

Pulmonary artery diastolic, mm Hg - - - 10 - 20 11 --

Central venous pressure, mm Hg - - - 4 - 11 13 --

Right ventricular stroke work index, g/m2 - - - 366 - 416 87 -

Mixed venous O2 saturation, % - - - 67 - 27 61 --

LVAD monitor

Revolutions per minute -- - 9000 6000 9000 6000 -- --

Flow, L/min -- - 5.3 "---" 6.6 "---" -- --

Pulsatility index -- - 5.4 6.2 3.7 6.3 -- --

Exercise test

VO2, ml/(kg × min) -- -- -- -- -- 15.3 -- 15.5

VE/VCO2 -- -- -- -- -- 75 -- 35

Respiratory exchange ratio -- -- -- -- -- 1.01 -- 0.9

Laboratory

Creatinine, μmol/L 64 109 61 - - - 124 95

NT-proBNP, pg/mL 2060 - 137 - - - 6530 312

Troponin T, ng/mL < 0.01 - < 0.01 - - - 0.90 -

Echocardiography

Left ventricular ejection fraction, % 10-15 70-75 75-80 70-75 75 70 60-70 55-60

Left ventricular end-diastolic diameter, mm - 42 37-40 39 42 37-40 35 51

Fractional shortening, % - 40 50 40 48 45 37 31

Right ventricular fractional area change, % - 35 38 40 - 39 - 32

Right ventricular end-diastolic diameter, mm - 28 27 27 23 22 27 29

Tricuspid regurgitation, grade, 1-4 - 2 < 1 < 1 - < 1 2 2

Tricuspid annular plane systolic excursion, mm - 8.5 9 9.5 - 10 11 15

E/A ratio - 2.2 1.6 2.8 - 1.8 1.6 2.5

E/E' ratio - - 12 15 - 11 - 18

Left atrial diameter, mm - 25 31 31 32 30 23 46

Mitral regurgitation, grade, 1-4 3 1 1 1 - 1 1 3

Septum during systole - right right left right left left left

Aortic valve opening 1/1 1/1 1/1 1/1 1/1 1/1 1/1 1/1

- not available

-- not applicable

"---" when flow is low, the HeartMate II controller does not estimate flow and the monitor instead displays "---"