MRI protocol in retinoblastomaa

Requirements
Scanner and coils
 Field strength above 1 T
 1.5-T system combined with one or two small surface coils (diameter < 5 cm)
 3.0-T system combined with multichannel head coil
Sequences (minimum requirements)
Orbits
 Transaxial T2-W (slice thickness ≤ 2 mm)
 Optional: Transaxial CISS (Siemens) / FIESTA (GE) / DRIVE (Philips)
Eye(s) and optic nerve(s)
 In-plane pixel size < 0.5 × 0.5 mm; slice thickness ≤ 2 mm
 Unilateral disease (or bilateral disease with only one eye strongly affected)
 Precontrast T1-W; at least one plane: transaxial or sagittal oblique
 T2-W; at least one plane: transaxial or sagittal oblique
 Postcontrast T1-W, no FS; transaxial and sagittal oblique
 Bilateral disease (both eyes strongly affected)
 Precontrast T1-W (transaxial)
 T2-W (transaxial)
 Postcontrast T1-W, no FS; sagittal oblique of both eyes and transaxial
Brain
 Transaxial T2-W (slice thickness ≤ 4 mm)
 Postcontrast T1-W (2D SE with slice thickness ≤ 3 mm or 3D GRE ≤ 1 mm)

FS fat-saturation, SE spin-echo, GRE gradient-echo

a Consensus among members of the European Retinoblastoma Imaging Collaboration (ERIC)