Table 7.
Use | Single-pulse | Paired-pulse, rTMS ≤1 Hz | Conventional rTMS, >1 Hz | Patterned rTMS (TBS, QPS) | Conventional or patterned rTMS with parameters exceeding those of Tables 3–5 |
---|---|---|---|---|---|
Research (Class 3 studies; Class 2 studies in normal subjects) | □ | □ | □ | □ ○ | ■ ○ |
Diagnostic applications | ■ ● | ■ ● | ■ ● | Currently not done | Currently not done |
Therapy/neuromodulation (Class 1 studies and Class 2 studies in patients) | Currently not done | ■ ● | ■ ● | ■ ● | ■ ● |
□ = Non medical setting allowable (i.e., psychology labs, robotics labs, research institutions, etc.). Presence of a Physician in the premises may not be required.
■ = Medical setting required
● = A medically responsible physician is required. Personnel skilled in the management of syncope and seizure is required in the lab.
○ = Emergency medical assistance is strongly recommended. Personnel skilled in the management of syncope and seizure should be present in the lab.