Skip to main content
. 2011 Dec 20;39(1):214–223. doi: 10.1118/1.3665767

Table 2.

Quantitative assessment of CT number variations in the phantom study before and after the use of TSFF.

  CT number range (HU) Standard deviation of PSR artifacts (HU) Noise (HU)
Location FS PS PS + TSFF FS PS PS + TSFF FS PS PS + TSFF
ROI 1 2.3 22.3 1.6 0.6 6.3 0.4 12.4 12.9 12.1
ROI 2 2.9 8.7 2.2 0.7 2.7 0.5 12.1 12.3 11.8
ROI 3 2.2 10.6 1.5 0.5 2.9 0.3 11.5 12.5 11.4