Comparison of glucose concentrations measured using the Optium Omega glucose meter and the Biosen C_line glucose analyzer. Passing-Bablok agreement test, slope: 0.91 (95% CI: 0.88 to 0.94), intercept: 6.5 (95% CI: 3.3 to 9.3), n = 289.