Table 1.

Results of the accuracy of the Helping Hand™ and the MEMS.

HH without reminder HH with reminder MEMS 18 month battery MEMS 2 year battery
Event level
Missing registrations 5/420 (1.20%) 1/630 (0.16%) 33/1050 (3.14%) 0/1050 (0%)
Overregistrations 2/420 (0.48%) 1/630 (0.16%) 0/1050 (0%) 0/1050 (0%)
Device level
Missing registrations 3/10 (30%) 1/15 (6.67%) 5/25 (20%) 0/25 (0%)
Overregistrations 2/10 (20%) 1/15 (6.67%) 0/25 (0%) 0/25 (0%)
100% correct functioning 7/10 (70%) 13/15 (87%) 20/25 (80%) 25/25 (100%)