Skip to main content
. Author manuscript; available in PMC: 2013 Feb 1.
Published in final edited form as: J Magn Reson Imaging. 2011 Oct 10;35(2):370–378. doi: 10.1002/jmri.22834

Table 1.

Intra-reader reproducibility errors for cartilage T2 measurements were calculated for each compartment and layer as the root mean square error (ms) and as the root mean square error coefficient of variation (%). LF: lateral femur, LT: lateral tibia, MF: medial femur, MT: medial tibia, and PAT: patella. The reproducibility errors averaged over all five compartments are printed in bold.

whole
compartment [%]
whole
compartment[ms]
deep layer
[%]
deep layer
[ms]
superficial
layer [%]
superficial
layer [ms]
LF 1.52 0.53 2.13 0.72 2.05 0.75
LT 0.90 0.28 2.08 0.61 1.57 0.55
MF 0.90 0.35 1.48 0.56 1.52 0.60
MT 2.61 0.88 2.40 0.81 4.46 1.67
PAT 2.38 0.71 4.92 1.38 1.63 0.55
mean 1.66 0.55 2.60 0.82 2.24 0.82