Skip to main content
. Author manuscript; available in PMC: 2012 Feb 3.
Published in final edited form as: Imaging Med. 2009 Oct;1(1):65–84. doi: 10.2217/iim.09.5

Table 3.

Summary of the four most common automatic exposure control strategies.

Manufa cturer AEC trade name Image quality reference Goal
General Electric Auto mA, Smart mA Noise index Maintain a constant noise level (defined in noise index), using tube
currents within prescribed minimum and maximum values
Toshiba SureExposure Standard deviation (high
quality, standard, low dose)
Maintain a constant noise level (defined in standard deviation
values for each protocol), using tube currents within preset
minimum and maximum values
Siemens CARE Dose4D Quality reference mAs Maintain the same image quality (varying noise target for different
attenuation level) with reference to a target effective mAs level for
a standard- sized patient
Philips DoseRight Reference image Keep the same image quality as in the reference image, regardless
of attenuation level

AEC: Automatic exposure control; mAs: Product of tube current and rotation time.