Table 1.

Specificity of the fiber-optic sensor for S. Enteritidis when compared with other live microorganisms.

Bacterial Culture Avg (pA)a Control (pA) Results
Salmonella Enteritidis 997.97 ± 91.85 285.08 Positive
Salmonella Typhimurium 868.13 ± 31.29 299.06 Positive
Escherichia coli O157:H7 570.00 ± 227.20 207.78 Negative
Staphylococcus aureus 906.39 ± 74.93 262.98 Positive
Enterococcus faecalis 789.87 ± 168.62 453.62 Negative
Lactobacillus rhamnosus 481.10 ± 166.96 311.07 Negative
Listeria monocytogenes 420.38 ± 95.86 270.60 Negative
a

Values are average of six optical fibers from three independent experiments with bacterial concentration of 5 × 108 cfu/mL and background (control) signal values, i.e., fibers exposed to buffer only are presented in separate column. Positives values were assigned based on Tukey’s grouping at P < 0.05.