Table 2.

Intra- and inter-assay imprecision of the LC-MS/MS assay. Imprecision was determined three times over the course of four days using ten replicates at each concentration.

Sample Conc (μM) from Nephelometry Intra-assay Conc (μM), Mean Intra-assay Imprecision (%) Inter-assay Conc (μM), mean Inter-assay Imprecision (%)
Patient 1 48.8 66.1 4.3 63.6 3.1
Patient 2 29.2 37.9 3.4 37.9 1.4
Patient 3 4.6 6.4 4.1 6.3 2.6