Skip to main content
. 2012 Mar;53(3):587–598. doi: 10.1194/jlr.D019745

TABLE 2.

Validation data for the analyzed retinoids

Analytes Equation for Calibration Curve r2 LLOQ (intraday, interday CV%)a Middle LOQ (intraday, interday CV%)b
atRA Y = 1.67x − 0.12 0.995 5.4, 15.0 3.6, 6.9
9-cisRA Y = 1.43x − 0.18 0.996 14.3, 19.6 2.1, 4.9
13-cisRA Y = 1.96x + 0.18 0.994 7.3, 13.7 3.2, 3.3
4oxo-atRA Y = 2.41x + 0.55 0.982 15.8, 17.9 4.0, 13.8
4oxo-13cisRA Y = 7.64x − 0.42 0.997 10.8, 18.8 6.5, 6.7
a

The LLOQs were determined at 0.05 nM RA isomers and 6 nM 4oxo-RA isomers.

b

The LOQs were determined at 7.5 nM RA isomers and 10 nM 4oxo-RA isomers.