Table 1.

Peak spot dose in cGy/MU at different depths for PPBSs of five different energies measured by ion chamber (IC) by placing it both at the peak of the single PPBS and the center of a 10 × 10 cm2 broad field. Spot spacing of 5 mm was used in creating the broad field. The nominal range corresponds to the depth of 90% Bragg peak dose location.

Energy (MeV)/nominal range (cm) Depth (cm) D0 from IC at peak of single spot (cGy/MU) D0 from IC in 10 × 10 cm2 field (cGy/MU) Overall uncertainty ±(%)
72.5/4.0 2 5.60 5.85 1.90
72.5/4.0 3 7.18 7.42 4.38
89.6/6.0 2 7.31 7.74 0.73
89.6/6.0 3 8.02 8.08 1.03
89.6/6.0 4 9.17 9.27 1.94
146.9/14.9 2 16.33 17.26 0.18
146.9/14.9 7 17.33 18.20 0.31
146.9/14.9 12 21.24 21.25 1.17
181.1/21.5 2 21.78 22.55 0.23
181.1/21.5 4 21.61 22.33 0.10
181.1/21.5 19 26.13 27.90 1.35
221.8/30.6 2 31.58 33.83 0.15
221.8/30.6 10 29.52 31.24 0.10
221.8/30.6 20 24.74 26.05 0.14