Published in final edited form as: IEEE Trans Med Imaging. 2011 Jun 20;30(11):1965–1973. doi: 10.1109/TMI.2011.2160192

Fig. 6.

Fig. 6

Results of the SNR comparison between the CMDM coil and the two-pole coil using a water phantom at 7T: (a) 1H image of CMDM coil; (b) 1H image of the two-pole coil; (c) SNR plot along the central line of the two images. The highest achievable SNR of the CMDM coil is 13% higher than that of the two-pole coil.