Skip to main content
. Author manuscript; available in PMC: 2013 Sep 1.
Published in final edited form as: Int J Radiat Oncol Biol Phys. 2012 Feb 11;84(1):130–136. doi: 10.1016/j.ijrobp.2011.11.041

Table 2.

3-Dimensional Target Excursions during IMRT Delivery

Left-Right Superior-Inferior Anterior-Posterior
n % n % n %
5 mm Tracking Limit Exceeded ≥ 1 second (out of 638 fractions) 10 1.6% 176 28% 159 25%
5 mm Tracking Limit Exceeded ≥ 30 seconds (out of 638 fractions) 1 < 0.5% 44 7% 31 5%
Target Excursion ≥ 1 cm (out of 638 fractions) 0 0 20 3% 13 2%
Area Under Curve ≥ 1 (out of 487 fractions) 1 <0.5% 39 8% 30 6%