Published in final edited form as: Magn Reson Med. 2011 Aug 16;67(3):880–889. doi: 10.1002/mrm.23059

Table 1.

Results of the static field homogeneity and image artifact tests show that the MRI system remains in specification with the treadmill positioned and operating next to the MRI table.

Specification Measured
Low High Baseline Treadmill Present Treadmill Running


Static Field Bpp 0.00 3.00 1.58 1.64 1.57
Homogeneity (ppm) Brms 0.00 0.40 0.10 0.11 0.10

Image Artifact (%) Echo 1 0.000 2.000 0.322 0.259 0.380
Echo 2 0.000 3.000 0.406 0.377 0.515