Skip to main content
. Author manuscript; available in PMC: 2013 Mar 1.
Published in final edited form as: Magn Reson Med. 2011 Aug 16;67(3):880–889. doi: 10.1002/mrm.23059

Table 3.

Treadmill performance testing inside and outside of the MRI room. Values represent average ± standard deviation across all 5 subjects. Only minimal differences in speed and elevation performance were observed between operation outside and inside the MRI room.

Desired Outside MRI Room Inside MRI Room Inside vs. Outside
Speed Elevation Speed Elevation Speed Elevation Speed Elevation
Actual Error Actual Error Actual Error Actual Error Difference
Stage (MPH) (deg) (MPH) (%) (deg) (%) (MPH) (%) (deg) (%) (%)
1 1.7 5.7 1.7 ± 0.0 2.2 5.8 ± 0.1 1.5 1.8 ± 0.0 4.1 5.8 ± 0.1 2.3 1.9 0.8
2 2.5 6.8 2.5 ± 0.1 2.2 6.9 ± 0.0 1.3 2.5 ± 0.1 2.3 6.9 ± 0.2 1.6 2.4 -0.5
3 3.4 8.0 3.4 ± 0.1 1.8 8.0 ± 0.1 1.0 3.4 ± 0.1 1.7 8.1 ± 0.1 0.9 1.3 0.7
4 4.2 9.1 4.1 ± 0.1 1.7 9.2 ± 0.1 1.0 4.2 ± 0.1 1.6 9.2 ± 0.1 1.2 1.8 0.0
5 5.0 10.2 5.0 ± 0.1 1.1 10.2 ± 0.1 0.6 5.0 ± 0.1 1.6 10.2 ± 0.1 1.0 0.6 -0.1
6 5.5 11.3 5.4 ± 0.1 1.6 11.0 ± 0.0 2.6 5.5 ± 0.1 1.7 11.0 ± 0.1 2.8 1.6 -0.2