. 2011 Aug 26;36(3):485–490. doi: 10.1007/s00264-011-1334-3

Table 2.

RSA results: head penetration (linear wear) at ten years in mm (95% CI), cup movements as translations in mm (95% CI) and rotations in degrees (95% CI). Precisions as mean difference between double examinations and level of significance

Wear results Cobalt-chrome Alumina Precision (SD) Significance
Proximal (y-axis) 0.96 (0.68–1.23) 0.42 (0.30–0.53) 0.08 (0.1) P < 0.001
3D (x + y + z-axis) 1.07 (0.79–1.35) 0.53 (0.38–0.63) P < 0.001
Subgroup SHP SP2 Cemex Palacos
Proximal (y-axis) 1.10 (0.74–1.47) 0.78 (0.31–1.25) 0.40 (0.23–0.57) 0.43 (0.25–0.62)
3D (x + y + z-axis) 1.17 (0.80–1.53) 0.96 (0.44–1.47) 0.49 (0.32–0.65) 0.52 (0.33–0.72)
Cup movements
Translation x-axis 0.43 (0.17–0.69) 0.28 (0.17–0.40) 0.04 (0.05) P = 0.19
Translation y-axis 0.71 (0.27–1.14) 0.33 (0.22–0.44) 0.03 (0.06) P = 0.63
Translation z-axis 0.35 (0.20–0.51) 0.27 (0.16–0.39) 0.08 (0.13) P = 0.32
Rotation x-axis 0.82 (0.36–1.29) 0.46 (0.27–0.64) 0.28 (0.74) P = 0.05
Rotation y-axis 0.83 (0.34–1.32) 0.47 (0.24–0.70) 0.28 (0.46) P = 0.06
Rotation z-axis 1.43 (0.18–2.68) 0.66 (0.41–0.92) 0.15 (0.40) P = 0.36

RSA radiostereometry, CI confidence interval, SD standard deviation