Skip to main content
. 2009 Jun 3;9(6):4309–4322. doi: 10.3390/s90604309

Table 2.

Response characteristics of the sulpiride-selective electrode.

Slope (mV per dec) ± S.D. 57.5 ± 0.7
Linear range (M) 5 × 10-5 to 1 × 10-2
Detection limit (M) ± S.D. 3.7 × 10-5± 7.0 × 10-6
Response time (s) 10-6 – 10-2 (M) t95% ≤ 15
Working pH range (10-6 – 10-2 M) 2 – 8
Lifetime (day) ≥ 15