. 2010 Apr 30;10(5):4430–4455. doi: 10.3390/s100504430

Table 2.

The expanded uncertainties (k = 2) of WTW OXI340i with a CellOx 325 sensor for laboratory and field conditions by model-based measurement uncertainty estimation.

tmeas (°C) Cmeas (mg/L) U (mg/L) U relative (%) U (mg/L) U relative (%)
Case 1 laboratory conditions Case 4 field conditions
20 0 0.10 - 0.10 -
5 12.71 0.24 1.9% 0.66 5.2%
15 10.01 0.10 1.0% 0.50 5.0%
20 9.01 0.07 0.8% 0.44 4.9%
25 8.18 0.08 0.9% 0.41 5.0%