Table 5.

The uncertainties of WTW OXI340i with a CellOx 325 sensor by the Nordtest approach.

tmeas (°C) Cmeas (mg dm−3) RMSbias (mg dm−3) u(bias) (mg dm−3) U (mg dm−3) U (%)

20 0 0.12 0.12 0.24
5 12.71 0.38 0.39 0.82 6.4
15 10.01 0.20 0.21 0.46 4.6
20 9.01 0.22 0.23 0.50 5.5
25 8.18 0.19 0.20 0.43 5.3