Tab. 5.

System suitability parameters and robustness

Robustness parameter Resolution between MH imp-2 and MH Tailing factor % RSD for 5 replicates

SP MH SP MH
Buffer pH 5.30 2.4 1.1 1.1 0.41 0.50
Buffer pH 5.50 2.3 1.0 1.1 0.32 0.63
Flow rate 0.18 mL/min 2.2 1.0 1.1 0.33 0.42
Flow rate 0.22 mL/min 2.3 1.1 1.0 0.21 0.30
Column temperature 20°C 2.1 1.1 1.0 0.20 0.31
Column temperature 30°C 2.2 1.0 1.1 0.34 0.43