Tab. 5.
System suitability parameters and robustness
Robustness parameter | Resolution between MH imp-2 and MH | Tailing factor | % RSD for 5 replicates | ||
---|---|---|---|---|---|
| |||||
SP | MH | SP | MH | ||
Buffer pH 5.30 | 2.4 | 1.1 | 1.1 | 0.41 | 0.50 |
Buffer pH 5.50 | 2.3 | 1.0 | 1.1 | 0.32 | 0.63 |
Flow rate 0.18 mL/min | 2.2 | 1.0 | 1.1 | 0.33 | 0.42 |
Flow rate 0.22 mL/min | 2.3 | 1.1 | 1.0 | 0.21 | 0.30 |
Column temperature 20°C | 2.1 | 1.1 | 1.0 | 0.20 | 0.31 |
Column temperature 30°C | 2.2 | 1.0 | 1.1 | 0.34 | 0.43 |