TABLE 3.

Accuracy of MDCT in Assessing In-Stent Restenosis by Stent Diameter

Reader Stent Size
< 2.5 mm (n = 20) 2.5–3.0 mm (n = 40) > 3.0 mm (n = 15)

Reader 1
% assessable segmentsa 45 58 62
Sensitivity (%) 50 (12–88) 50 (12–88) 100
Specificity (%) 80 (31–97) 74 (52–88) 57 (20–87)
Reader 2
% assessable segmentsa 45 53 46
Sensitivity (%) 25 (3–76) 25 (3–76) 100
Specificity (%) 80 (31–97) 88 (65–97) 40 (10–80)

Note—Data in parentheses are 95% CI.

a

A chi-square test with 2 degrees of freedom was performed to evaluate the difference between assessable and nonassessable segments across stent diameter categories. A statistical difference in evaluability was not observed in either reader 1 (p = 0.590) or reader 2 (p = 0.838).