Published in final edited form as: Ultrasound Med Biol. 2012 Feb 15;38(4):626–641. doi: 10.1016/j.ultrasmedbio.2012.01.004

Figure 4.

Figure 4

(A) Empirical receiver-operating characteristic (ROC) curves for accuracy of locally predicted tissue ablation over all specimens (n = 12) for each analyzed parameter (dashed lines) and its temporal extremum (solid lines). The curves are generated by plotting the true-positive pixel fraction TPFtot versus the false-positive FPFtot of each parameter. The ROC curves allow the relative assessment of each parameter’s ability to classify the lesion. Table 1 lists the corresponding area under the curve (AUC) for each curve shown. (B) Pixel fraction identified as a function of the frame-to-frame decorrelation parameter for pixels in the pre-HIFU region, including all specimens. Because no bubble activity should exist prior to HIFU exposure, this graph was used to propose a threshold for identification of bubble activity of σtot0=1.22, where the pixel fraction goes to zero.