Table 4.

Figures of merit: comparison of performance of the FIA and MCFA flow systems.

Parameter FIA system MCFA system
Detection limit (LD)
(3σ/slope of calibration, n = 10)
0.59 μg L−1 in solution 0.08 μg L−1 in solution
Quantification limit (LQ)
(10σ/slope of calibration, n = 10)
1.95 μg L−1 in solution 0.27 μg L−1 in solution
Linearity (μg · L−1) 2–50 (r 2 = 0.999) 0.27–27 (r 2 = 0.999)
Precision s r (%) (n = 5) <10 <10
Sampling frequency (hour−1) 70 160
Sample consumption per determination (mL) 0.5 0.6
Reagent consumption for each determination 0.25 mL HCl*
NaBH4: 7.3 mg
*concentrated
0.05 mL HCl*
NaBH4: 2.5 mg
*concentrated