Table 4.
Figures of merit: comparison of performance of the FIA and MCFA flow systems.
| Parameter | FIA system | MCFA system |
|---|---|---|
| Detection limit (LD) (3σ/slope of calibration, n = 10) |
0.59 μg L−1 in solution | 0.08 μg L−1 in solution |
| Quantification limit (LQ) (10σ/slope of calibration, n = 10) |
1.95 μg L−1 in solution | 0.27 μg L−1 in solution |
| Linearity (μg · L−1) | 2–50 (r 2 = 0.999) | 0.27–27 (r 2 = 0.999) |
| Precision s r (%) (n = 5) | <10 | <10 |
| Sampling frequency (hour−1) | 70 | 160 |
| Sample consumption per determination (mL) | 0.5 | 0.6 |
| Reagent consumption for each determination | 0.25 mL HCl* NaBH4: 7.3 mg *concentrated |
0.05 mL HCl* NaBH4: 2.5 mg *concentrated |