Electrospray ionization-tandem mass spectrometry operating parameters for bile acid intermediates

Glycine conjugates (IS included) Taurine conjugates Sulfatides Glucuronides
Nebulizer gas (au) 11 11 11 11
Curtain gas (au) 9 9 9 9
Collision gas (au) 5 5 5 5
Ion spray voltage (V) -4100 -4100 -4100 -4100
Temperature (°C) 200 200 200 200
Declustering potential (V) -35 -35 -35 -35
Focusing potential (V) -31 -31 -31 -31
Entrance potential (V) -10 -10 -10 -10
Prefilter (V) 23 23 23 23
Collision energy (V) -55 -129 -60 -75
Collision cell exit potential (V) -9 -9 -9 -9
Daughter ion (m/z) 74 80 97 85
Scan range of precursor ions (m/z) 430 to 550 490 to 590 460 to 490 600 to 630

The mass spectrometer was operated in a negative precursor ion scan mode with a step-size of 0.1 amu. Nitrogen was used as nebulizer, curtain and collision gases. au: Arbitrary units; IS: D4-glycocholic acid.