Skip to main content
. 2011 Dec 2;26(3):454–462. doi: 10.1038/eye.2011.293

Table 3. Summary of descriptive statistics and paired-sample t-test of the central subfield (C) thickness of Stratus and Spectralis OCT measurements, and Spectralis-predicted values in eyes with DME.

Descriptive statistics
  N Minimum Maximum Mean SE SD
CStratus (measured thickness) 130 247 849 426 11.39 129.82
CSpectralis (measured thickness) 130 289 903 487 11.15 127.11
CSpectralis (matched locations) 130 304 903 488 11.13 126.92
CSpectralis (predicted values) 130 327 946 511 11.72 133.60
Paired sample t-test
  Paired correlation
Paired differences
  N Correlation Sig. Mean SD SE CI95 of difference
              Lower Upper Sig.
CStratus vs CSpectralis (measured thickness) 130 0.987 0.000 –61 21.23 1.86 –64 –57 0.000
CStratus vs CSpectralis (matched locations) 130 0.994 0.000 –62 14.75 1.29 –65 –60 0.000
CSpectralis (matched locations) vs CSpectralis (predicted values) 130 0.987 0.000 –24 22.33 1.96 –28 –20 0.000
CSpectralis (matched locations) vs CSpectralis (predicted values) 130 0.994 0.000 –23 16.12 1.41 –25 –20 0.000

Abbreviations: CI95, 95% confidence interval; N, number; SE, standard error; SD, standard deviation.