Table 2.

Table showing mean differences in terms of mean Axial Length, Anterior Chamber Depth, Keratometry and Intraocular Lens Power and limits of agreement between biometry devices.

Mean differences and limits of agreement (LoA) between biometry devices
Axial length (mm) LoA ACD LoA Keratometry (D) LoA IOL Power (D) LoA
Jasvinder et al.[23] LCOR-PCI 0.01 ± 0.03 −0.04 to 0.07 N/A N/A −0.11 ± 0.18 −0.45 to 0.59 0.07 ± 0.26 −0.44 to 0.24
Hoffer et al.[21]LCOR-PCI* 0.026 −0.05 to 0.10 0.128 −0.12 to 0.38 −0.107 −0.46 to 0.25
Hoffer et al.[21]LCOR-PCI** 0.023 −0.05 to 0.099 0.146 −0.33 to 0.63 −0.121 −0.54 to 0.30
Salouti et al.[22]US-PCI −0.01±0.09 −0.18 to 0.16 0.07±0.19 −0.29 to 0.44 −0.05±0.27 (Javal vs. PCI) −0.58 to 0.48 (Javal vs. PCI) −0.01±0.33 (SRK/T) −0.66 to 0.64 (SRK/T)
Salouti et al.[22]US-LCOR 0.002±0.07 −0.14 to 0.14 −0.03±0.24 −0.50 to 0.44 0.61±0.30 (Javal vs. LCOR) 0.01 to 1.20 (Javal vs. LCOR) 0.10±0.40 (SRK/T) −0.68 to 0.88 (SRK/T)
Salouti et al.[22]PCI-LCOR 0.01±0.07 −0.12 to 0.14 −0.10±0.26 −0.60 to 0.40 0.65±0.27 (PCI vs. LCOR) 0.07 to 1.20 (PCI vs. LCOR) 0.11±0.40 (SRK/T) −0.67 to 0.89 (SRK/T)
Chenat al.[20] 0.01±0.05 −0.09 to 0.11 N/A N/A −0.08±0.15 N/A N/A N/A
*

Cataract eyes,

**

Clear lenses

PCI: Partial coherence interferometry, LCOR: Low coherence optical reflectometry

LoA: 95% limits of agreement

ACD: Anterior chamber depth