Skip to main content
. Author manuscript; available in PMC: 2013 May 1.
Published in final edited form as: Optom Vis Sci. 2012 May;89(5):E652–E666. doi: 10.1097/OPX.0b013e318238c34e

Table 1.

Instrument algorithm RNFL thickness measures for quadrant and clock hours for Cirrus and Spectralis SD-OCT. Along with the paired t-test statistic, the intraclass correlation coefficient (ICC) with 95% confidence intervals provides a measure of agreement for the two segmentation algorithms.

Cirrus Spectralis 95% Confidence Paired
t-test
Mean
(µm)
Std Dev Mean
(µm)
Std Dev ICC Lower Upper
Global 91.17 8.42 97.92 7.94 0.623 −0.074 0.864 <0.001
Temporal 63.71 12.96 72.79 12.71 0.765 −0.054 0.938 <0.001
Superior 115.03 14.94 118.52 12.98 0.734 0.552 0.847 0.022
Nasal 65.89 10.06 74.56 15.88 0.559 0.119 0.780 <0.001
Inferior 120.08 16.39 125.21 15.21 0.764 0.544 0.876 0.001
Sector 1 49.57 8.70 58.81 10.87 0.631 −0.071 0.887 <0.001
Sector 2 77.15 19.18 85.73 17.53 0.853 0.068 0.957 <0.001
Sector 3 125.78 21.95 134.40 23.92 0.780 0.491 0.895 <0.001
Sector 4 114.12 24.88 112.97 22.18 0.797 0.658 0.883 0.613
Sector 5 105.17 16.77 108.18 18.31 0.794 0.653 0.882 0.075
Sector 6 80.23 17.45 90.59 23.26 0.743 0.256 0.893 <0.001
Sector 7 55.31 9.79 62.95 15.50 0.414 0.110 0.640 <0.001
Sector 8 62.12 10.60 70.13 16.12 0.571 0.182 0.776 <0.001
Sector 9 95.75 18.21 99.21 18.00 0.701 0.517 0.824 0.100
Sector 10 134.94 26.79 137.62 27.59 0.801 0.666 0.885 0.301
Sector 11 129.53 26.72 138.45 25.62 0.885 0.465 0.959 <0.001
Sector 12 64.40 15.79 73.37 15.84 0.759 0.080 0.915 <0.001