Published in final edited form as: Optom Vis Sci. 2012 May;89(5):E652–E666. doi: 10.1097/OPX.0b013e318238c34e

Table 3.

Instrument algorithm vs. Custom RNFL thickness measures for quadrant and clock hours for Cirrus and Spectralis SD-OCT.

Cirrus Spectralis Paired t-
test
Instrument
Global
Thickness
(µm)
Custom
Global
Thickness
(µm)
ICC Paired
t-test
Instrument
Global
Thickness
(µm)
Custom
Global
Thickness
(µm)
ICC
Global 91 ± 8 110 ± 8 0.227 <0.001 98 ± 8 110 ± 8 0.405 <0.001
Temporal 64 ± 13 76 ± 10 0.515 <0.001 73 ± 13 77 ± 10 0.803 <0.001
Superior 115 ± 15 133 ± 14 0.407 <0.001 119 ± 13 135 ± 15 0.463 <0.001
Nasal 66 ± 10 88 ± 13 0.249 <0.001 75 ± 16 89 ± 14 0.539 <0.001
Inferior 120 ± 16 143 ± 15 0.361 <0.001 125 ± 15 140 ± 15 0.600 <0.001
Sector 1 50 ± 9 59 ± 6 0.303 <0.001 59 ± 11 60 ± 6 0.542 0.520
Sector 2 77 ± 19 83 ± 13 0.588 0.005 86 ± 18 84 ± 12 0.692 0.231
Sector 3 126 ± 22 138 ± 27 0.636 <0.001 134 ± 24 138 ± 29 0.768 0.233
Sector 4 114 ± 25 140 ± 24 0.405 <0.001 113 ± 22 144 ± 21 0.404 <0.001
Sector 5 105 ± 17 119 ± 22 0.540 <0.001 108 ± 18 123 ± 23 0.594 <0.001
Sector 6 80 ± 17 110 ± 24 0.334 <0.001 91 ± 23 112 ± 24 0.570 <0.001
Sector 7 55 ± 10 78 ± 16 0.204 <0.001 63 ± 15 77 ± 15 0.486 <0.001
Sector 8 62 ± 11 77 ± 11 0.304 <0.001 70 ± 16 76 ± 12 0.593 0.001
Sector 9 96 ± 18 111 ± 16 0.469 <0.001 99 ± 18 111 ± 17 0.684 <0.001
Sector 10 135 ± 27 152 ± 25 0.603 <0.001 138 ± 28 147 ± 25 0.836 <0.001
Sector 11 130 ± 27 166 ± 23 0.375 <0.001 138 ± 26 162 ± 21 0.525 <0.001
Sector 12 64 ± 16 84 ± 15 0.420 <0.001 73 ± 16 86 ± 16 0.600 <0.001