Table 1.

Assay performance parameters at representative points along the dynamic range.

[1,25(OH)2D3] (ng/mL) Accuracy
(%)
Precision
(%)
Recovery
(%)
Plasma
0.1 95.0 2.5 -
1.0 110.7 1.9 92.2 (11.4)
10 97.4 0.4 -
100 87.3 2.1 -
Tumor homogenate
0.1 110.7 2.9 -
0.3 105.0 2.6 -
1.0 89.6 0.8 11.0 (16.1)
3.0 96.3 1.2 -

Assay performance was assessed in triplicate using samples that contained the indicated concentrations of 1,25(OH)2D3. Precision was calculated as the relative standard deviation. Accuracy was calculated as the back-calculated concentration relative to the nominal concentration. Extraction recovery was assessed at 1 ng/mL 1,25(OH)2D3 and was calculated as the relative response of an extracted sample versus the response of a standard in mobile phase.