Skip to main content
. 2012 Jan 17;9:2. doi: 10.1186/1743-0003-9-2

Table 3.

Agreement limits for the sensors embedded in the glove and traditional goniometry

Finger joint Mean difference (bias) SD of differences 95% agreement limit
Index DIP -0.43 1.94 -4.31 to 3.44
Index PIP -0.46 1.95 -4.36 to 3.44
Index MCP -0.48 2.18 -4.84 to 3.89
Thumb IP -0.98 2.59 -6.15 to 4.20
Mean -0.59 -4.91 to 3.74