Patients |
1. The sensor device must be tested under real-life conditions, similar to its use in clinical practice (in order to take into account most random interferences). |
|
2. The sensor device must be tested in a population of the intended use (i.e., similar patient group as target patient population, e.g., diabetes of pregnancy vs TGC in critically ill patients). |
|
3. Sufficient number of patients and sufficient number of measurements per patient are required for statistical reasons. |
|
Blood glucose range |
4. Glucose measurements should be spread over the full glycemic range. |
|
5. The median and interquartile ranges of the glucose measurements must be reported to clarify the clinical context and compared with the patient's target glucose range. |
|
Glucose sensing in the ICU |
6. Sampling of arterial blood is a prerequisite when applying TGC in the ICU. In noncritically ill patient populations, the use of venous/capillary blood is accepted, assuming physiological features (e.g., lag time) typical of the sampling compartment are understood. |
|
Evaluation methodology |
7. Glucose sensor performance should be evaluated both for the full glycemic range as well as for the individual hypo/normo/hyperglycemic range. |
|
8. Limits of agreement (e.g., Bland-Altman analysis) should be smaller than the difference of the patient's target zone (e.g., 110 mg/dl ? 80 mg/dl = 30 mg/dl in the context of TGC in critically ill patients), preferably over the entire glycemic range, in case of nonpersistent measurement behavior for each individual hypo/normo/hyperglycemic range. |
|
9. Sensor accuracy should be computed with respect to clinically defined criteria, preferably statistically based (e.g., GLYCENSIT Phase 2 for the ISO criteria). |
|
10. Overestimation measurement behavior in the hypoglycemic range may lead to clinically wrong treatment decisions and should be avoided accordingly. |