. 2012 Apr 16;4(4):99–107. doi: 10.4253/wjge.v4.i4.99

Table 1.

Results of PillCam colon trials

n S (%) E (%) PPV (%) NPV (%)
Results for polyps (any size)
Eliakim et al[8] 2006 91 69 81 74 78
Schoofs et al[9] 2006 41 76 64 83 54
Van Gossum et al[21] 2009 320 72 78 - -
Eliakim et al[10] 20091 98 88 - - -
Results for significant polyps (≥ 6 mm or > 3 polyps > 3 mm)
Eliakim et al[8] 2006 91 63 94 67 91
Schoofs et al[9] 2006 41 60 73 46 83
Van Gossum et al[21] 2009 320 64 84 - -
Eliakim et al[10] 20091 98 89 76 - -
Spada et al[24] 20111 109 84 64 - -
Results for polyps ≥ 10 mm
Eliakim et al[10] 20091 98 88 89 - -
Spada et al[24] 20111 109 88 95 - -
Results for other lesions - - - - -
Schoofs et al[9] 2006 41 76 63 82 52
Eliakim et al[8] 2006 91 78 76 47 93
Van Gossum et al[21] 2009 320 76 - - -
1

The second generation of PillCam colon capsule endoscopy. S: Sensitivity; E: Specificity; PPV: Positive predictive value; NPV: Negative predictive value.