Skip to main content
. Author manuscript; available in PMC: 2012 Apr 20.
Published in final edited form as: Nat Protoc. 2009 Nov 5;4(12):1722–1736. doi: 10.1038/nprot.2009.183

Table 2.

Troubleshooting table.

Step Problem Possible reason Solution
11 Low chromosome yield Low mitotic index Optimize incubation time with metaphase-
arresting drug
Undissociated chromosome clusters Extend vortexing time or syringe the suspension to
release the chromosomes into suspension
18 High percentage of debris Check pH of hypotonic solution and
polyamine isolation buffer
Replace reagents (Hypotonic and polyamine
isolation buffer)
Chromosome fragmentation due to
physical force
Reduce vortexing time
Poor data resolution Insufficient staining time Extend fluorescence dyes staining time
Effect from sodium sulfite and sodium
citrate
Omit either reagent
30 Low yield after GenomePlex WGA Derivative chromosome DNA is
degraded or was considerably < 10 ng
Increase the amount of input DNA
Fragmentation time was not optimal A 4-min fragmentation time is crucial
Negative control produced a product Reagents contaminated Replace reagents
47 Flow through is brightly colored
and the membrane has retained
only a small amount of color
Cleanup column membrane failed Retrieve the sample by putting flow through back
into a new column and clean up as described in
the protocol
54 Low yield and low SA after labeling Klenow inactive Test reagents
83 No fluorescent signal on microarray Check the slide was in the correct
orientation during hybridization
Repeat experiment
Low signal Poor atmospheric conditions Monitor the ozone levels, temperature and humidity
in the laboratory where the experiment was conducted
and, in particular, in the proximity of the scanner