Table 2.
Troubleshooting table.
Step | Problem | Possible reason | Solution |
---|---|---|---|
11 | Low chromosome yield | Low mitotic index | Optimize incubation time with metaphase- arresting drug |
Undissociated chromosome clusters | Extend vortexing time or syringe the suspension to release the chromosomes into suspension |
||
18 | High percentage of debris | Check pH of hypotonic solution and polyamine isolation buffer |
Replace reagents (Hypotonic and polyamine isolation buffer) |
Chromosome fragmentation due to physical force |
Reduce vortexing time | ||
Poor data resolution | Insufficient staining time | Extend fluorescence dyes staining time | |
Effect from sodium sulfite and sodium citrate |
Omit either reagent | ||
30 | Low yield after GenomePlex WGA | Derivative chromosome DNA is degraded or was considerably < 10 ng |
Increase the amount of input DNA |
Fragmentation time was not optimal | A 4-min fragmentation time is crucial | ||
Negative control produced a product | Reagents contaminated | Replace reagents | |
47 | Flow through is brightly colored and the membrane has retained only a small amount of color |
Cleanup column membrane failed | Retrieve the sample by putting flow through back into a new column and clean up as described in the protocol |
54 | Low yield and low SA after labeling | Klenow inactive | Test reagents |
83 | No fluorescent signal on microarray | Check the slide was in the correct orientation during hybridization |
Repeat experiment |
Low signal | Poor atmospheric conditions | Monitor the ozone levels, temperature and humidity in the laboratory where the experiment was conducted and, in particular, in the proximity of the scanner |