. 2012 Mar 29;6:503–510. doi: 10.2147/OPTH.S29865

Table 3.

Correlation matrix showing the correlation between ECLoss% (percentage of postoperative endothelial cell loss) at 3 months and different phaco parameters

Age CDE Torsional T. Volume Phaco T. Average Phaco Power Average Phaco Power/3 Aspiration T.
Spearman’s rho ECLoss% Correlation coefficient −0.077 0.425 0.334 0.278 0.345 0.307 0.343 0.176
Sig. (2-tailed) 0.406 0.000 0.000 0.002 0.000 0.001 0.000 0.054
N 120 120 120 120 120 120 120 120

Note: Spearman’s rank-order correlation coefficient, rho, (ρ) was used to evaluate the relationship because at least one of the continuous variables is not normally distributed.

Abbreviations: ECLoss%, percentage of endothelial cell loss; US Tot. Eq. P/3, ultrasonic total equivalent power in position 3; Avg_Phaco_Power/3, Average phaco power in position 3; Eq. Avg. Tor. Ampl/3, Equivalent average torsional amplitude in position 3; CDE, cumulative dissipated energy. Torsional T, torsional time, Aspiratio T, Aspiration time; Phaco T, phaco time.