Table 2. Precision of duplicate testing for all 3 methods.
Test Type | Number of pairs | MeanDifference | 95% CI | P-value* |
Pima Capillary | 130 | 33.38 | 14.19, 52.57 | 0.005 |
Venous | 204 | −6.54 | −13.69, 0.61 | 0.099 |
BD FACS Venous | 206 | 1.87 | −6.77, 10.51 | 0.899 |
Using Wilcoxon Signed Rank Test;