Skip to main content
. 2012 Mar 22;2012:101249. doi: 10.1155/2012/101249

Table 1.

Concentration data form validation.

Spiked plasma concentration (pg/ml) Concentration measured (pg/ml) Mean ± Sd
n = 5
Precision (% CV) Accuracy %
50.00 49.92 ± 0.44 0.80 99.23
100.00 100.21 ± 1.91 1.90 98.14
500.00 502.73 ± 6.83 1.40 98.65
1000.00 1000.54 ± 11.53 1.10 98.92
5000.00 5005.06 ± 38.75 0.80 99.28
10000.00 9978.95 ± 160.56 1.60 98.45
20000.00 19871.04 ± 303.46 1.50 98.53
30000.00 29759.82 ± 508.47 1.70 98.37
40000.00 40310.88 ± 123.85 0.30 99.74
50000.00 50084.85 ± 266.72 0.50 99.55