Skip to main content
. 2012 Mar 22;2012:101249. doi: 10.1155/2012/101249

Table 2.

Precision and accuracy (analysis with spiked plasma samples at three different concentrations).

Spiked plasma concentration (pg/ml) Within-run (n = 6) Between-run (n = 30)
Concentration measured (n = 6) (pg/ml) (mean ± Sd.) Precision (%CV) Accuracy % Concentration measured (n = 30) (pg/ml) (mean ± Sd.) Precision (%CV) Accuracy %
150.00 143.40 ± 3.20 2.20 95.60 143.50 ± 3.90 2.70 95.70
15000.00 14746.40 ± 338.40 2.30 98.30 14719.10 ± 248.30 1.70 98.10
35000.00 34935.20 ± 730.40 2.10 99.80 34699.20 ± 498.40 1.40 99.10