Skip to main content
. 2012 Mar 22;2012:101249. doi: 10.1155/2012/101249

Table 3.

Stability of Memantine in human plasma samples.

Spiked plasma concentration (pg/ml n = 6) Concentration measured (pg/ml) Precision (%CV) Accuracy (%)
Room temperature stability for 26 hr in plasma
150.00 151.23 1.00 100.82
35000.00 34926.50 0.80 99.79
Three freeze-thaw cycles
150.00 153.02 2.30 102.01
35000.00 34818.00 0.90 99.48
Auto sampler stability for 79 hr
150.00 154.59 1.40 103.06
35000.00 35213.50 0.90 100.61
Stability for 76 days −30°C
150.00 151.53 3.20 101.02
35000.00 35080.50 1.40 100.23