Table 1.

Results of method validation for monacolin K (lovastatin) using 1H NMR

Parameter Result
Investigated linear working range 5 - 100 mg/L
Limit of detectiona 6 mg/L (0.33 mg/capsulec)
Limit of quantitationa 13 mg/L (0.72 mg/capsulec)
Intraday precisionb
 Authentic sampled 10%
 Standard solution 7.3%
Interday precisionb
 Authentic sampled 11%
 Standard solution 8.2%
Recovery 104% (at 50 mg/L)
113% (at 20 mg/L)

a Limit of detection (LOD) and quantitation were determined by establishing a separate calibration curve near LOD. The limits were calculated from the residual standard deviation of the regression line [42]

b Precisions were expressed as relative standard deviation (%), intraday (n = 5), interday (n = 10)

c Based on a sample weight of 0.9 g (2 capsules) in 50 mL ethanol

d Total statins in authentic samples