Skip to main content
. 2011 Dec 24;52(13):9658–9664. doi: 10.1167/iovs.11-8559

Table 5.

The Failure Criteria and their Corresponding Sensitivities at the Various Specificity Levels

Hyperopia Myopia Astigmatism Anisometropia Group 1 Conditions Any VIP-Targeted Eye Conditions Specificity for Any VIP-Targeted Eye Conditions
NCR
2.00 1.00 0.50 1.00 0.98 0.88 0.50
1.75 0.25 0.75 0.75 0.96 0.84 0.60
2.00 0.25 1.00 1.00 0.96 0.81 0.70
2.50 2.00 1.00 1.00 0.96 0.76 0.80
2.50 1.25 1.25 1.00 0.92 0.71 0.85
2.75 2.75 1.25 1.50 0.90 0.64 0.90
2.50 2.00 2.00 2.00 0.85 0.56 0.95
Retinomax
1.50 2.00 0.75 0.75 0.96 0.90 0.50
0.75 2.50 0.75 2.25 0.96 0.88 0.60
0.75 2.50 1.00 2.25 0.95 0.83 0.70
1.25 4.00 1.00 1.75 0.92 0.77 0.80
1.00 3.75 1.25 2.25 0.91 0.73 0.85
1.75 3.75 1.25 2.75 0.87 0.68 0.90
1.75 4.00 1.75 2.75 0.83 0.58 0.95
SureSight
2.25 0.50 0.75 1.00 0.98 0.91 0.50
3.25 0.75 0.75 1.75 0.95 0.88 0.60
2.50 0.75 1.25 1.25 0.95 0.83 0.70
3.25 0.75 1.25 2.25 0.90 0.77 0.80
4.25 0.75 1.25 3.00 0.87 0.72 0.85
3.75 0.75 1.75 2.75 0.82 0.65 0.90
5.00 1.00 2.00 4.00 0.77 0.55 0.95