Skip to main content
. 2012 Mar 21;7:30. doi: 10.1186/1746-1596-7-30

Table 1.

Hemoglobin level measured using HemoCue® venous sample, HemoCue® capillary sample, and automated hematology analyzer (g/dl)

Method of measurement Mean ± SD Median (min; max)
HemoCue® (venous) 12.70 ± 1.77 12.80 (8.90; 17.90)
HemoCue® (capillary) 12.87 ± 2.04 12.90 (8.20; 17.70)
Hematology analyzer 11.53 ± 1.63 11.40 (8.10; 15.00)