Table 2.

Correlation, bias, and limits of agreement between hemoglobin level using HemoCue® and automated hematology analyzer

Comparison of methods Correlation Coefficient Bias ± SD (95%CI) Limits of agreement
HemoCue® (venous) vs. Hematology analyzer 0.58 1.17 ± 1.57 (0.87; 1.47) -1.97 to 4.31
HemoCue® (capillary) vs. Hematology analyzer 0.51 1.34 ± 1.85 (0.99; 1.69) -2.36 to 5.04
HemoCue® (venous) vs. HemoCue® (capillary) 0.51 -0.17 ± 1.90 (-0.53; 0.19) -3.97 to 3.63